| |
A
common problem in water reclamation plants
is selecting proper instrumentation equipment.
Plant personnel will try to choose equipment
that monitors accurately with minimal
maintenance. Users see equipment accuracy
presented in many ways, however, and it’s
important to understand the differences.
|
If an instrument has an accuracy claim of 0.5%
of full scale, for instance, you should recognize
that the actual accuracy diminishes as the operating
conditions fall below the full-scale setting.
Sometimes, though, you will hear claims that
a meter offers 0.5% of reading over full range.
Although the difference may sound insignificant,
it could be very costly to the owner. Imagine,
for instance, that a paddle-wheel flowmeter
claims to have an accuracy of ±0.5%.
Suppose, further, it is a percent of full range,
and the full range is 50 feet per second (ft/sec).
If the flow range you will use it in is 6
ft/sec, which is common in treatment plants,
the actual accuracy is much different than you
might expect: 0.005 x 50 ft/sec = ±0.25
ft/sec .If you apply this accuracy against a
flow rate of 6 ft/sec, you see that the actual
accuracy is: ±0.25/6 ft/sec = ±0.0417,
or 4.17% . Comparing a magnetic flowmeter with
an accuracy of 0.5% of reading to a Doppler
flowmeter with an accuracy of 0.5% of full range
yields a similar result.
A common problem occurs when a city or municipality
uses two different types of flowmeters. Imagine
one meter is a highly accurate magnetic flowmeter
located in a meter vault to monitor the plant’s
effluent flow, and the other is a Doppler meter
monitoring the influent flow; this meter’s
accuracy diminishes as the flow ate drops. Case
histories have shown that the plant appears
to be either generating wastewater, because
the effluent is more than the influent, or something
is evaporating the wastewater.
We know in both cases that neither of these
conditions really exists. What is really happening
is that the Doppler meter is not matching the
accuracy of the magnetic meter. The difference
between 0.5% of 12 million gallons a day (Mgd)
and 4.17% of 12 Mgd is substantial: 4.17% –
0.5% x 12 Mgd = 0.44 Mgd, or 305 gal/min .Matters
are made even worse if the Doppler meter is
used for pacing chemical feed into the wastewater
with the same inaccuracies, resulting in either
overdosing or underdosing.
Water treatment plants have low, average daily,
and high peak demand flows, and further, low
and average daily flows occur more frequently.
This demonstrates the importance of being cautious
in choosing meter types for those flow variables.
Many types of flowmeters suffer in performance
as the flows decrease and approach the lower
end of their viable flow range; therefore, pacing
during low flow periods may be highly suspect.
Chemicals are becoming more costly, analytical
instruments for measuring the effects of these
chemicals are becoming costly, and corrosion
due to underdosing or overdosing wastewater
can be costly to equipment. All of these may
contribute to effluent that is a danger to wildlife
and, in extended cases, human life.
Repeatability
Another tool in evaluating equipment is repeatability,
defined as the quantity that characterizes the
ability of an instrument to give identical indications
or responses for repeated applications of the
same value of the quantity measured under the
same conditions of use. In the past, when equipment
operated on motion balance, where equipment
used linkages and temperature compensation values,
repeatability was critical. Today, however,
a number of field instruments work on force
balance techniques, such as piezoelectric crystals,
capacitance, and strain gauges.
These all work on the principle that if you
put a force on an instrument, there should be
no motion, though an electric signal is generated
on the output of that instrument. There are
still flow, level, and chemical measuring devices
that do not work on the force balance principle,
and for these types, looking at the repeatability
of that piece of equipment is still important.
A steady widening of the repeatability is an
indication that something is going wrong with
the instrument.
Although some might believe good repeatability
is a measure of accuracy, that is incorrect.
To understand the difference between accuracy
and repeatability, imagine an archer shooting
at a conventional archery target. Suppose one
archer hits the bulls-eye consistently. Because
he was always accurate, the shots were repeatable.
Now imagine an archer that hits the target but
misses the bulls-eye consistently. Although
the archer has good repeatability, the archer
was not accurate. Good repeatability does not
guarantee accuracy. If you do not see a proper
accuracy statement on equipment but only a repeatability
statement, be cautious.
Rangeability and uncertainty
One of the most common problems with instrumentation
equipment is the exaggeration of its range.
How many times have you heard a meter can read
flow rates at velocities of 1–100 ft/sec,
giving the impression that you can read flows
accurately through that total velocity range?
What often goes unmentioned is that the particular
meter’s accuracy has a 10:1 turndown ratio.
This means that a meter sized to measure a range
of 0–30 Mgd has a true accuracy over the
full range 3–30 Mgd. Below 3 Mgd, the
meter accuracy diminishes. Additionally, different
types of meters have different turndown ratios
over their full range.
It is common for a Venturi tube, for example,
to have two transmitters measuring the flow.
This is because a Venturi tube with one transmitter
measures accurately with a 6:1 turndown ratio
over the full range. So if we look at a range
of 0–30Mgd, the meter’s accuracy
diminishes below 5 Mgd. The range over which
the instrument meets the stated linearity of
uncertainty requirements is its “rangeability.”
Uncertainty is the range of values within which
the true value lies with a specified probability.
Uncertainty of ±1% at 95% confidence
means the instrument will give the user a range
of ±1% for 95 readings out of 100.
Another common error occurs during the equipment
sizing. In the water reclamation industry, it
is a common practice to assume that solids in
wastewater will settle out around a velocity
of 2 ft/sec. A magnetic flowmeter reads accurately
if the minimum velocity is above 2 ft/ sec,
but below this, settling is likely to occur—
and who can then say what the accuracy really
is?
Typically, designers size plants to handle
increased flow capacities for 20 years. For
this reason, designers often oversize pipes
for early life-cycle flow, and there is corresponding
settlement inside the pipe. This settling can
also occur in the inner liner of the meters.
Because these meters are velocity sensing devices
with an assumed constant cross section, they
will give a false reading if the inner liner
becomes coated with sludge. A solution may be
to reduce the size of the meter to increase
velocity by utilizing a pipe reducer on the
inlet side and a pipe expansion section on the
discharge side of the meter.
If possible, avoid connecting the reducer
and expander directly onto the meter. Manufacturers
recommend that when you reduce the pipe, the
flowmeter has a minimum of six to 10 pipe diameters
upstream from an elbow or valve and at least
two pipe diameters downstream of a pipe elbow
or valve. This provides a less distorted flow
profile for the meter to read. Be certain you
can afford to lose the pressure head when you
reduce the meter. Maximum velocities should
not exceed 15 ft/sec. By maintaining a minimum
scouring effect inside the pipe, your sludge
buildup inside pipes and any in-line equipment
will diminish, helping avoid measurement errors
and costly maintenance downtime.
Misconceptions and truths
Some people will ask for the accuracy of a
certain flowmeter, level, or pressure-measuring
device and, upon hearing a low number, think
that everything involved with the flowmeter
will be of the same accuracy. However, the meter
accuracy is not the accuracy for the entire
flow system. A mathematical equation known as
the root mean square (RMS) correctly determines
the accuracy of the complete system.
Consider the case of a magnetic flowmeter that
records flow locally, sending an analog signal
to an operator’s workstation via a programmable
logic controller (PLC). You must look at each
component’s accuracy: a magnetic flowmeter
(±0.5%), a magnetic flowmeter transmitter
(±0.5%), a wire connection to the recorder
(±0.01%), a wire connection to a local
control panel terminal block (±0.01%),and
the I/O card of the PLC (0.4%). Each component
in the system has its own measurement errors
and uncertainties, which contribute to the overall
accuracy of the complete system. In real cases,
there could be more components attached to a
control system. To use the RMS method, first
square each number, yielding 0.000025, 0.000025,0.00000001,
0.00000001, and 0.000016.
Second, add the numbers. Then find the square
root of the sum. The entire system has an accuracy
of approximately ±0.813% instead of 0.5%.
This accuracy equation works for any individual
chemical, pressure, level, temperature, or flow
loop. Remember, too, that no two flowmeters
or instruments will have exactly the same accuracy.
For this reason, the accuracy statement should
indicate a ± component. When choosing
an instrumentation control strategy, look at
all the manufacturers’ equipment literature
regarding accuracy; consider the range, repeatability,
turndown ratio, and piping constraints; choose
similar equipment types; and utilize the RMS
equation.
This article was written and submitted by
John Davis is a senior member of ISA who operates
Davis
Instrumentation Services. His address is nira47@aol.com.
|