HomeContact usSite MapLinks+Other info
Unit ConverterCompaniesSubmitted ArticlesFeedback
Technical Glossary
A | B | C | D | E | F | G | H | I | J | K | L | M | N | O | P | Q | R | S | T | U | V | W | X | Y | Z | #
Z

Zero Adjustments:
Used when 'setting up' a transducer to adjust the output signal to zero when zero load/pressure is applied.

Zero Balance: The output signal of the transducer with rated excitation and with no-load applied, usually expressed as a percent of rated output.

Zero Offset: 1. The difference expressed in degrees between true zero and an indication given by a measuring instrument. 2. See Zero Suppression

Zero Point: The electrical zero point where zero millivolts would be displayed. Used in conjunction with the slope control to provide a narrower range calibration.

Zero Power Resistance: The resistance of a thermistor or RTD element with no power being dissipated.

Zero Return: The difference in zero balance measured immediately before rated load application of specified duration and measured after removal of the load, and when the output has stabilized.

Zero Suppression: The span of an indicator or chart recorder may be offset from zero (zero suppressed) such that neither limit of the span will be zero. For example, a temperature recorder which records a 100° span from 400° to 500° is said to have 400° zero suppression.

Zero Voltage Switching: The making or breaking of circuit timed such that the transition occurs when the voltage wave form crosses zero voltage; typically only found in solid state switching devices.



Copyright © 2002 Flowmeterdirectory.com